GET THE APP

..

Biosensors & Bioelectronics

ISSN: 2155-6210

Open Access

Immunosensor Characterization Using Impedance Spectroscopy

Abstract

Lin K, Wai Kwan Lim, New Siu Yee, Su Xiao Di and Liu Bin

Non-invasive and high sensitive electrochemical impedance spectroscopy (EIS) has been used for qualitative characterization of stepwise surface preparation process for constructing an immunosensor, starting from plain silicon substrate, 3-aminopropyltriethoxysilane (APTES) silanization, glutaraldehyde functionalization and analytes (rat immunoglobulin G immobilization, followed by protein G-coated gold nanoparticles), using both the horizontal direction (2 terminal) and vertical direction (3 terminal) characterization setup. The investigation demonstrated that 2-terminal setup is able to review the quality of antibody and antigen molecules bonding and distinguish the effects from the underlayers on the immuosensors; while 3-terminal setup is able to correlate the thickness and dielectric constant of each layers. It correlated well to the incubation time of preparing different sensors as well. The two different EIS measurement setups are able to serve different needs for immunosensor characterization and EIS measurement shows higher sensitivities compared with UV-Vis absorption measurement.

PDF

Share this article

Google Scholar citation report
Citations: 6207

Biosensors & Bioelectronics received 6207 citations as per Google Scholar report

Biosensors & Bioelectronics peer review process verified at publons

Indexed In

 
arrow_upward arrow_upward