The Calibration Process of OSL Detectors used for Staff and Patient Dosimetry in Hospital Environment
Author(s): Nina Tuncel, Binnur Karayalcin and Gulsah kocaNina Tuncel, Binnur Karayalcin and Gulsah koca
The optically stimulated luminescence (OSL-BeO) dosemeter is increasingly being used as a dosimetric technique in various fields such as medical dosimetry. According to our fixed dose protocol, the activities of 3.7-7.4 GBq I-131 source is used for thyroid carcinoma therapy. The in house calibration process for usage of OSL’s at hospital was arranged according to the encapsulated I-131. The measurement point was planned in three different radial distances from source free in air. The dose-rate measurement was done by Geiger-Muller (GM), and then three pieces of OSL was placed in the same positions for one hour. The inverse square law consistency was found (R2=0.99). The calibration coefficient was calculated. For determining the performance of OSL at different dose rates, it used for personnel and patient dosimetry. The average annual dose/2mon to the whole body for all staff by.. Read More»