The results of a TLD Therapy Dosimetry Quality Assurance Program for Dong Nai General Hospital, Vietnam (Period 2013-2017)
Author(s): Duong Thanh Tai and Truong Thiet DungDuong Thanh Tai and Truong Thiet Dung
The objective of this study is to perform a thermo luminescent dosimetry (TLD) postal dose audit provided the
International Atomic Energy Agency (IAEA) for a linear accelerator in Dong Nai General Hospital. The TLD were
irradiated with an absorbed dose to water of 2 Gy for 6 MV and 15 MV photon beams generated by a linear
accelerator (LINAC) (Siemens Medical Solutions, Concord, CA). The percentage deviation relative to IAEA
measured dose was 0.4% and 3.4% for 6 MV and 16 MV, respectively. Agreement within ± 5% between the users
stated dose and the IAEA measured dose is considered satisfactory. The results showed that all photon beams are
within the IAEA’s acceptance limit of ± 5%. As a conclusion, the beam output of Siemens Primus is accuracy in
delivery of absorbed dose within ± 5%... Read More»